User Manual
Table Of Contents
- Table of contents
- Glossary
- Before you begin...
- The basics...
- Basic programming
- Starting on insulin
- Using Bolus Wizard
- Optimizing pump therapy
- Insulin pump therapy follow-up
- Utilities
- Troubleshooting and alarms
- Troubleshooting
- My pump has a no delivery alarm...
- What happens if I leave the battery out too long?
- Why doesn’t my pump battery last very long?
- What is a CHECK SETTINGS alarm?
- My screen appears distorted...
- I can’t get out of the priming loop...
- The pump is asking me to rewind...
- My bolus stopped...
- My pump buttons are not acting right during a bolus...
- My pump won’t display my BG reading from my meter...
- I dropped my pump
- I submerged my pump in water
- Alarms
- Alarm conditions
- Troubleshooting
- Pump maintenance
- User safety
- Pump specifications
- Alarms and error messages
- Alarm history
- Backlight
- Basal
- BG target
- Bolus delivery
- Bolus history
- Bolus units
- Bolus Wizard
- Carb ratios
- Carb units
- Daily totals
- Default screen
- Delivery accuracy
- Drive motor
- Dual Wave bolus
- Easy bolus
- Infusion pressure
- (insulin) sensitivity
- Insulin type
- Low resv (reservoir) warning
- Meter value
- Normal bolus
- Occlusion detection
- Percent temp basal
- Power supply
- Prime function
- Prime history
- Program safety checks
- Pump size
- Pump weight
- Remote control
- Reservoir
- Square Wave bolus
- Status screen
- Temporary (temp) basal rate
- Time and date screen
- Water tight
- Bolus Wizard specifications
- Default settings
- Icon table
- Menu map
Starting on insulin 39
Rewinding your pump
Before you continue, make sure the pump is NOT connected to your body.
1. If you removed your reservoir and are replacing it, go to the REWIND screen.
Main > Prime > Rewind
NOTE - If you just changed your insulin type and the HOME screen
appears, you can press
ACT
to make the REWIND screen appear.
2. In the REWIND screen, press ACT to start the rewind process.
The REWINDING screen will appear while the pump rewinds.
3. After the pump rewinds, the MANUAL PRIME screen will
appear.
NOTE - If you press
ESC
, the pump will go back to the REWIND
screen. Once you start a rewind, you cannot cancel it.
If you are practicing, do these steps:
a. Do NOT insert the reservoir in your pump.
Make sure the shipping cap is installed in the reservoir
compartment.
b. Continue with the manual prime instructions described in the “Manual prime” section on page 41.
If you are not practicing, continue to the next section, “Inserting the reservoir in your pump.”
WARNING: Make sure the infusion set is disconnected from your body before you rewind or prime the
pump. Never insert the reservoir into the pump while the tubing is connected to your
body. Doing so could result in an accidental infusion of insulin.
REWIND
Disconnect from body
Press ACT to Rewind
MiniMed
ACT
REWIND
Disconnect from body
Press ACT to Rewind
MiniMed MiniMed
REWINDING
Wait for notification
REWIND
Disconnect from body
Press ACT to Rewind
ACT
MiniMed
MiniMed
MANUAL PRIME
Disconnect from Body
Insert/Lock Reservoir
Hold ACT to Prime
MiniMed
ESC
MiniMed
ACT
time out
(HOME screen)