User Manual
Table Of Contents
- Table of contents
- Glossary
- Before you begin...
- The basics...
- Basic programming
- Starting on insulin
- Using Bolus Wizard
- Optimizing pump therapy
- Insulin pump therapy follow-up
- Utilities
- Troubleshooting and alarms
- Troubleshooting
- My pump has a no delivery alarm...
- What happens if I leave the battery out too long?
- Why doesn’t my pump battery last very long?
- What is a CHECK SETTINGS alarm?
- My screen appears distorted...
- I can’t get out of the priming loop...
- The pump is asking me to rewind...
- My bolus stopped...
- My pump buttons are not acting right during a bolus...
- My pump won’t display my BG reading from my meter...
- I dropped my pump
- I submerged my pump in water
- Alarms
- Alarm conditions
- Troubleshooting
- Pump maintenance
- User safety
- Pump specifications
- Alarms and error messages
- Alarm history
- Backlight
- Basal
- BG target
- Bolus delivery
- Bolus history
- Bolus units
- Bolus Wizard
- Carb ratios
- Carb units
- Daily totals
- Default screen
- Delivery accuracy
- Drive motor
- Dual Wave bolus
- Easy bolus
- Infusion pressure
- (insulin) sensitivity
- Insulin type
- Low resv (reservoir) warning
- Meter value
- Normal bolus
- Occlusion detection
- Percent temp basal
- Power supply
- Prime function
- Prime history
- Program safety checks
- Pump size
- Pump weight
- Remote control
- Reservoir
- Square Wave bolus
- Status screen
- Temporary (temp) basal rate
- Time and date screen
- Water tight
- Bolus Wizard specifications
- Default settings
- Icon table
- Menu map
Starting on insulin 37
5. With the vial up, slowly
pull down on the plunger
to fill the reservoir.
6. Tap the side of the reservoir
to make any air bubbles rise
to the top of the reservoir.
7. Slowly push up on the plunger
just enough to remove any air
bubbles from the reservoir.
Slowly pull down on the plunger
to completely fill the reservoir.
8. While holding the reservoir, turn it counter-
clockwise then pull straight up to remove it
from the transfer guard and vial.
9. Put the tubing connector on the reservoir.
Turn it clockwise to secure.
10. Tap the side of the reservoir
to remove any air bubbles.
11. To purge air bubbles that have
risen to the top of the reservoir,
push up on the plunger until you
see insulin in the tubing.
12. Without pulling, turn the plunger
counter-clockwise to remove it
from the reservoir.
“CLICK”