36 Chapter 4
Filling the reservoir
WARNING: Your pump will not accurately deliver insulin if air bubbles are in the reservoir or
the infusion set. To prevent this, take care to remove air bubbles when filling your
reservoir.
1. Remove reservoir from package. Make sure
plunger rod is fully extended.
2. Swab vial with alcohol.
Making sure you do not push down on the plunger,
press the transfer guard onto the vial.
3. Snap transfer guard into locked position. 4. Push down on the plunger to pressurize the vial.
plunger
reservoir
transfer
guard
rod
vial
locked
“CLICK”