24 Chapter 3
Maximum bolus limit
The maximum bolus (max bolus) is a safety feature that limits the amount of insulin that can be delivered in a single
bolus. The factory setting is 10.0 units. You can specify the limit from 0.0 to 25.0 units. It is important to discuss this
feature with your healthcare professional to determine your maximum bolus amount.
To set the maximum bolus limit, do these steps:
1. Go to the MAX BOLUS SETUP screen.
Main > Bolus > Max Bolus
2.
Set your maximum bolus limit
and press ACT.
3. Your maximum bolus
is set. Exit the menus.
Example #1: Max bolus
Shelby takes very small doses of insulin for her meal boluses. As a safety limit, she and her healthcare
professional reset her pump with a maximum bolus of 5.0 units.
Example #2: Max bolus
David is a growing teenager. He loves to eat big meals and requires very large doses of insulin for his food. He
reset his pump with a maximum bolus of 20.0 units so he can take more insulin when he needs to.
(flashing)
MiniMed
MAX BOLUS SETUP
10.0
U
0-25
x __
MAX BOLUS SETUP
15.0
U
0-25
ACT
MiniMed
BOLUS MENU
Use Bolus Wizard
Manual Bolus
Bolus History
ESC
MiniMed