130 Chapter 12
Default settings
Menu Item Default Setting Limits Increments Warning Limits
Bolus
Menu:
*Bolus wizard: Off
Easy bolus: On
Easy bolus step: 0.1 u/h maximum bolus setting
Dual/Square bolus: Off
Maximum bolus: 10.0 u/h 0 - 25 u (per single bolus)
BG reminder: Off 0:00 - 5:00 0:30 (minutes)
Basal
Menu:
Patterns: Off
Maximum basal rate: 2.0 u/h 0.00 - 35.00 u/h 0.05u
Basal rate: 0.0 u/h 0.05u
Temp basal type: U/H max basal rate setting 5U/H (or 5%) (basal rate = 0.0)
Prime
Menu:
Insulin type: U100 fast acting
Utilities
Menu:
(Alarm) History: (no defaults)
Alert type: audio, beep-med
Auto-off: Off
Low reservoir warning: (20) insulin units 5 - 50 u; 2nd: @ 1\2 amount
(2:00 - 24:00; 2nd: after 1:00)
20 u
(0:30 mins)
(Time/Date) Time: 12 a.m. (midnight)
(Time/Date) Date: 1/1/03
(Time/Date) Time format: 12-hour
Block: Off
Alarm clock: Off
Remote option: Off
Meter option: Off
Language: English
(*Bolus Wizard settings)
carb units: grams or
exchanges
0 - 300 g/u or
0 - 20.0 u/exch
1 g/u or
0.5 u/exch
(none)
ins to carb (or exch) ratio: 15 grams/u or
1 units/exch
3 - 75 g/u or
0.2 - 5.0 u/exch
1 g/u or
0.1 u/exch
5 - 50 g/u or
0.3 - 3.0 u/exch
(insulin) sensitivity: 50
mg/dL or
2
.8 mmol/L
10 - 250 mg/dL or
0.5 - 13.9 mmol/L
1 mg/dL or
0.1 mmol/L
20 - 100 mg/dL or
1.1 - 5.6 mmol/L
BG target: 100
mg/dL or
5
.6 mmol/L
80 - 160 mg/dL or
4.4 - 8.9 mmol/L
1 mg/dL or
0.1 mmol/L
90 - 140 mg/dL or
5.0 - 7.8 mmol/L