User Manual
Table Of Contents
- Table of contents
- Glossary
- Before you begin...
- The basics...
- Basic programming
- Starting on insulin
- Using Bolus Wizard
- Optimizing pump therapy
- Insulin pump therapy follow-up
- Utilities
- Troubleshooting and alarms
- Troubleshooting
- My pump has a no delivery alarm...
- What happens if I leave the battery out too long?
- Why doesn’t my pump battery last very long?
- What is a CHECK SETTINGS alarm?
- My screen appears distorted...
- I can’t get out of the priming loop...
- The pump is asking me to rewind...
- My bolus stopped...
- My pump buttons are not acting right during a bolus...
- My pump won’t display my BG reading from my meter...
- I dropped my pump
- I submerged my pump in water
- Alarms
- Alarm conditions
- Troubleshooting
- Pump maintenance
- User safety
- Pump specifications
- Alarms and error messages
- Alarm history
- Backlight
- Basal
- BG target
- Bolus delivery
- Bolus history
- Bolus units
- Bolus Wizard
- Carb ratios
- Carb units
- Daily totals
- Default screen
- Delivery accuracy
- Drive motor
- Dual Wave bolus
- Easy bolus
- Infusion pressure
- (insulin) sensitivity
- Insulin type
- Low resv (reservoir) warning
- Meter value
- Normal bolus
- Occlusion detection
- Percent temp basal
- Power supply
- Prime function
- Prime history
- Program safety checks
- Pump size
- Pump weight
- Remote control
- Reservoir
- Square Wave bolus
- Status screen
- Temporary (temp) basal rate
- Time and date screen
- Water tight
- Bolus Wizard specifications
- Default settings
- Icon table
- Menu map
128 Chapter 12
■
Active insulin only reduces the correction portion of the estimate, not the food portion.
■
If the active insulin is more than the correction estimate, the correction portion of the estimate will be changed to zero (0).
■
If the current BG is lower than target, the correction portion of the estimate will reduce the total estimate.
■
If a Dual Wave bolus is less than the estimate due to the max bolus limit or user change, the square (sq) portion is reduced first.
Bolus Wizard examples: (settings:) Carb ratio: 30 grams/unit
Insulin
sensitivity: 40 mg/dL/unit
Target BG: 120 mg/dL
#1: No previous bolus activity for at least 10 hours. User inputs 60 grams of carbs and does not enter BG.
#2: No previous bolus activity for at least 10 hours. User inputs 60 grams of carbs and a BG of 200.
#3: No previous bolus activity for at least 10 hours. User inputs 60 grams of carbs and a BG of 80.
60g
30g/u
(food estimate)
= 2 units
estimate = 2 units
60g
30g/u
(food estimate)
= 2 + 2
+
200mg/dL - 120mg/dL
40mg/dL/u
(correction)
= 2 units = 2 units
estimate = 4 units
= 4 units
60g
30g/u
(food estimate)
+
80mg/dL - 120mg/dL
40mg/dL/u
(correction)
= 2 units =
-40mg
40mg
= -1 unit
= 2 + (-1)
estimate = 1 unit
= 1 unit