User Manual

Table Of Contents
122 Chapter 12
Easy bolus
Features programming using audible tones (or vibrate pulses) in user-determined increments. Beep mode range: 0
to maximum bolus; Vibrate mode range: 0 to 20 steps or maximum bolus, whichever comes first.
Default step increment: 0.1 unit
Step size < maximum bolus.
Adjustable step size: 0.1 to 2.0 units per step.
Accessible from the remote control or pump buttons.
Infusion pressure
Maximum infusion pressure and occlusion pressure: 13.7 PSI.
(insulin) sensitivity
Maximum settings: 8
Factory default: 50 mg/dL (2.8 mmol/L)
Insulin type
Pump users can select one: U100 fast-acting (default) U100 regular-acting
Low resv (reservoir) warning
Values are based on displayed amount, not actual amount.
Meter value
The BG measurement received from Paradigm Link meter. Appears in the ENTER BG screen during bolus program-
ming. Appears on the screen when the pump is idle at the HOME screen.
Expiration: 12 minutes;
Range: 20 600 mg/dL
maximum meter entries: 3
range: 10 250 mg/dL
(0.5 13.9 mmol/L)
warning limits: less than 20 or greater than 100 mg/dL
(less than 1.1 or greater than 5.6 mmol/L)
time: 2 24-hours, and
@ 1:00 hour before empty
08:00 hours (default when time is selected)
units: 5 50 units, and
@ 1/2 amount remaining
20 units (factory default)