User Manual
Table Of Contents
- Table of contents
- Glossary
- Before you begin...
- The basics...
- Basic programming
- Starting on insulin
- Using Bolus Wizard
- Optimizing pump therapy
- Insulin pump therapy follow-up
- Utilities
- Troubleshooting and alarms
- Troubleshooting
- My pump has a no delivery alarm...
- What happens if I leave the battery out too long?
- Why doesn’t my pump battery last very long?
- What is a CHECK SETTINGS alarm?
- My screen appears distorted...
- I can’t get out of the priming loop...
- The pump is asking me to rewind...
- My bolus stopped...
- My pump buttons are not acting right during a bolus...
- My pump won’t display my BG reading from my meter...
- I dropped my pump
- I submerged my pump in water
- Alarms
- Alarm conditions
- Troubleshooting
- Pump maintenance
- User safety
- Pump specifications
- Alarms and error messages
- Alarm history
- Backlight
- Basal
- BG target
- Bolus delivery
- Bolus history
- Bolus units
- Bolus Wizard
- Carb ratios
- Carb units
- Daily totals
- Default screen
- Delivery accuracy
- Drive motor
- Dual Wave bolus
- Easy bolus
- Infusion pressure
- (insulin) sensitivity
- Insulin type
- Low resv (reservoir) warning
- Meter value
- Normal bolus
- Occlusion detection
- Percent temp basal
- Power supply
- Prime function
- Prime history
- Program safety checks
- Pump size
- Pump weight
- Remote control
- Reservoir
- Square Wave bolus
- Status screen
- Temporary (temp) basal rate
- Time and date screen
- Water tight
- Bolus Wizard specifications
- Default settings
- Icon table
- Menu map
119
Chapter 12
Pump specifications
This section provides detailed information on specifications related to your pump. The safety features of your
pump are individually listed and described.
Alarms and error messages
Indicators: audible tone (beep) or vibration (silent)
All alarms and errors show messages on the pump’s screen and provide instructions on what to do. Unresolved
alarms will escalate to siren mode for added safety.
Alarm history
Maximum records shown: 36
Backlight
LCD (Liquid Crystal Display) type
Time-out: 7 seconds (in the HOME screen);
30 seconds (after the last button press).
Basal
Delivery: 0.05 – 35 units/hour (Maximum units: 35/hour)
Factory maximum setting: 2.0 units/hour
Increments: 0.05 units
3 patterns maximum, each with 48 rates maximum
BG target
Maximum targets: 8
range: 80 – 160 mg/dL
(4.4 – 8.9 mmol/L)
warning limits: less than 90 or greater than 140 mg/dL
(less than 5.0 or greater than 7.8 mmol/L)