112 Chapter 9
Failed batt test
The pump tests the voltage of
each battery installed. This
test ensures a battery with
low voltage is not used. If the
battery does not have enough
voltage, this alarm will occur. The pump will not function
and the battery must be replaced. (Always make sure that
you install a NEW battery into the pump.)
Is priming complete?
If you manually prime your
pump with more than 30u insu-
lin, this message will appear.
Press ESC, ACT to clear the
message. If manual prime is
complete, press ESC. If the manual prime is not complete,
press and hold ACT until manual prime is complete.
Max delivery
This alarm alerts you when you
have taken more insulin than
expected based on maximum
bolus and maximum basal rates.
Motor error
Insulin delivery has stopped.
This alarm will occur if your
pump detects a motor error.
No delivery
Insulin delivery has stopped.
This alarm will occur if your
pump detects a blockage.
No reservoir
The reservoir is not inserted
correctly or no reservoir has
been inserted.
Off no power
The battery is dead. Replace
battery immediately. Follow
the directions on the screen.
Check to make sure that the
time is correct on the
screen. Reset the time if
necessary.
Reset
Reset alarm triggers when
pump settings are cleared
because:
pump was cleared
(clear pump function) and
settings have not been reprogrammed, or
download attempt from PC is incomplete.
(Download function is applicable to the optional
software feature. Refer to the software user guide
for more information.)
Weak Battery
(712 pump only)
The pump tests the voltage
of each battery installed. If
the battery voltage is less
than full strength, this alarm may occur. The pump will
function, but the battery life will be shorter than
expected.
Always make sure to install a NEW battery in the pump.
BATT TEST
ESC, ACT to clear
Replace battery now
8:35A
FAILED
Delivery stopped
Use 1 AAA alkaline
MiniMed
IS PRIMING
Disconnect from body
COMPLETE?
See user guide
for instruction
ESC, ACT to clear
MiniMed
3:25P
MAX
Exceeded 1 hour
max delivery
Check BG
ESC, ACT to clear
DELIVERY
MOTOR
12:05A
Delivery stopped
Disconnect set
ESC, ACT to clear
ERROR
MiniMed
11:17A
Delivery stopped
Change entire set
Check BG
See user guide
to troubleshoot
ESC, ACT to clear
DELIVERY
NO
MiniMed
NO
RESERVOIR
4:36P
Delivery Stopped
Change reservoir
ESC, ACT to clear
MiniMed
OFF
NO POWER
04:42
Delivery stopped
Use 1 AAA alkaline
ESC, ACT to clear
Replace battery now
0% battery life
MiniMed
RESET
Settings cleared
by user
Reprogram settings
ESC, ACT to clear
8:35A
11:17A
Shorter battery
life expected
ESC, ACT to clear
BATTERY
WEAK
MiniMed