User Manual
Table Of Contents
- Table of contents
- Glossary
- Before you begin...
- The basics...
- Basic programming
- Starting on insulin
- Using Bolus Wizard
- Optimizing pump therapy
- Insulin pump therapy follow-up
- Utilities
- Troubleshooting and alarms
- Troubleshooting
- My pump has a no delivery alarm...
- What happens if I leave the battery out too long?
- Why doesn’t my pump battery last very long?
- What is a CHECK SETTINGS alarm?
- My screen appears distorted...
- I can’t get out of the priming loop...
- The pump is asking me to rewind...
- My bolus stopped...
- My pump buttons are not acting right during a bolus...
- My pump won’t display my BG reading from my meter...
- I dropped my pump
- I submerged my pump in water
- Alarms
- Alarm conditions
- Troubleshooting
- Pump maintenance
- User safety
- Pump specifications
- Alarms and error messages
- Alarm history
- Backlight
- Basal
- BG target
- Bolus delivery
- Bolus history
- Bolus units
- Bolus Wizard
- Carb ratios
- Carb units
- Daily totals
- Default screen
- Delivery accuracy
- Drive motor
- Dual Wave bolus
- Easy bolus
- Infusion pressure
- (insulin) sensitivity
- Insulin type
- Low resv (reservoir) warning
- Meter value
- Normal bolus
- Occlusion detection
- Percent temp basal
- Power supply
- Prime function
- Prime history
- Program safety checks
- Pump size
- Pump weight
- Remote control
- Reservoir
- Square Wave bolus
- Status screen
- Temporary (temp) basal rate
- Time and date screen
- Water tight
- Bolus Wizard specifications
- Default settings
- Icon table
- Menu map
Utilities 101
Clear pump
The clear pump function resets the time and date and clears all pump settings except for language. The pump
settings are restored to the factory defaults. When you clear your pump, you must reprogram all your settings
before you can use your pump again. The pump does not clear the internal pump memory.
Do these steps to clear your pump:
1. Go to the CLEAR PUMP screen. Select Settings and press ACT.
Main > Utilities > Clear Pump
2.
The CONFIRM screen will ask you to verify that you want to clear the
pump. Select Yes then press ACT.
3. The RESET message will appear indicating the pump settings have been
cleared. Read the instructions on the screen then press ESC, ACT to
clear the message.
4. The pump will go through various screens while it restarts.
After the pump clears all your settings, the screen will go to the
TIME/DATE SETUP screen.
5. Reset the time and date as described in the section, “Setting the time and date” in chapter 3.
6. After you set the time and date, you must rewind your pump. Refer to the section, “Rewinding your pump” in
chapter 4 for instructions. Remember, all your setting have been cleared and you must reprogram all your
settings.
WARNING: Do not clear your pump while it is connected to your body.
CAUTION: Do not clear your pump unless directed by your healthcare professional or a Medtronic MiniMed
representative. If you clear your pump, it will be necessary to reprogram all your personal pump
settings as directed by your healthcare professional. Additionally, you will have to rewind your
pump.
MiniMed
CLEAR PUMP
Settings
ACT
MiniMed
CONFIRM
Yes
No
ACT
14:05
RESET
Cleared settings
Program as required
ESC, ACT to clear
MiniMed