Utilities 97
Turn on remote control option
1. Go to the REMOTE OPTION screen. Select On and
press ACT.
Main > Utilities > Remote Options
2.
The REMOTE ID MENU screen will appear. Add, delete or
review your remote ID as described in the next section.
Exit the menus if you are done.
Add, delete, review remote control IDs
Each remote control has its own unique ID. Up to three (3) different remote control IDs can be programmed in your
pump. The remote control programming screens are very similar to those for the meter. Make sure to select
“Remote Options” (in the UTILITIES MENU) when programming your remote control.
If you are not sure that your remote control ID is entered in your pump, check the REVIEW REMOTE ID screen.
You have to turn on the remote option to add, delete or review the remote control ID(s) programmed in your
pump.
1. In the REMOTE OPTION screen, select On and press ACT.
The REMOTE ID MENU will appear.
2. Add, delete or review your remote ID(s) as desired.
NOTE - The remote control RF ID code is on the back of
the remote control.
WARNING: If there is “Low Battery” condition, the pump will not receive signals from the remote. To
ensure the pump communicates with the remote control, make sure the pump does not have
a low battery.
(Replacing the low battery with a new battery will restore remote control function.)