User Manual

Table Of Contents
Utilities 95
NOTE - Be aware. If you use “time” as the low resv warning type and you deliver large boluses, the actual
time remaining could be less than the warning time. “Time” low resv warning types are intended to let
you know if you will have enough insulin while you are sleeping.
Review daily insulin totals
The DAILY TOTALS screen provides a day-by day history of the total amount of insulin that delivered for the past
14 days. This screen includes all bolus and basal amounts delivered midnight-to midnight for each of the past 14
days. The first line in the DAILY TOTALS screen shows the amount of insulin you delivered so far that day.
NOTE - The insulin used to prime your pump are not included in the DAILY TOTALS screen. This
amount is counted separately and shown in the PRIME HISTORY screen.
Why should I review my daily totals?
Comparing your daily insulin deliveries to your blood glucose records helps you and your healthcare professional
identify your optimal daily insulin rate(s).
What is included in the daily totals?
Daily totals include all basal and bolus insulin deliveries, but they do not include insulin used for priming your
pump. Each total reflects all basal and bolus insulin delivered for that day.
Where is the daily totals screen?
Main > Utilities > Daily Totals
For “Insulin units”:
Enter the number of units you want remaining
when the first warning will go off. Press ACT.
The pump will alarm first when the specified
units remain, then again when half that remaining
amount is used.
For “Time”:
Enter the amount of remaining time you want for the
first warning. Press ACT.
The pump will alarm first when the specified time
remains, then again one hour before empty.
MiniMed
WARNING UNITS
_ _ U
(flashing)
MiniMed
WARNING TIME
_ _ : _ _ Hours
(flashing)
DAILY TOTALS
Today
25 FEB
MiniMed
26 FEB
26.5U
54.5U
48.5U
date
(total) basal and bolus deliveries