86 Chapter 6
Selecting temp basal type
Your pump will remember the temp basal type setting. Once the type is set, you do not have to set it again. To
select a temp basal type, do these steps:
1. Go to the BASAL MENU.
Main > Basal > Temp Basal Type
Select Temp Basal Type and press ACT.
2. The SET TEMP BASAL AS screen will appear.
Select Insulin Rate or Percent of Basal and press ACT.
3. The screen will return to the BASAL MENU. The temp basal type is now set.
Exit the menus.
NOTE - Be aware. If your temp basal type is set to “Percent of Basal,”
changes to your basal rate are not allowed until after temp basal is
completed or canceled.
Deliver a temp basal
NOTE - A temp basal cannot exceed your programmed max basal rate.
1. Go to the BASAL MENU.
Main > Basal > Set/Edit Temp Basal
Select Set/Edit Temp Basal and press ACT.
2. The SET DURATION screen will appear. The duration* will flash. Enter the desired
minutes or hours (30 minutes - 24-hours), then press ACT.
NOTE - * Duration is the amount of time it will take for the pump to
deliver the temporary basal.
Temp Basal Type
Max Basal Rate
Patterns
BASAL MENU
MiniMed
ACT
MiniMed
SET TEMP BASAL AS
Insulin Rate U/H
Percent of Basal
ACT
BASAL MENU
Set/Edit Temp Basal
Set/Edit Basal
Basal Review
MiniMed
ESC
BASAL MENU
Set/Edit Temp Basal
Set/Edit Basal
Basal Review
MiniMed
ACT
MiniMed
SET DURATION
3:00
Temp
ACT