User Manual
Table Of Contents
- Table of contents
- Glossary
- Before you begin...
- The basics...
- Basic programming
- Starting on insulin
- Using Bolus Wizard
- Optimizing pump therapy
- Insulin pump therapy follow-up
- Utilities
- Troubleshooting and alarms
- Troubleshooting
- My pump has a no delivery alarm...
- What happens if I leave the battery out too long?
- Why doesn’t my pump battery last very long?
- What is a CHECK SETTINGS alarm?
- My screen appears distorted...
- I can’t get out of the priming loop...
- The pump is asking me to rewind...
- My bolus stopped...
- My pump buttons are not acting right during a bolus...
- My pump won’t display my BG reading from my meter...
- I dropped my pump
- I submerged my pump in water
- Alarms
- Alarm conditions
- Troubleshooting
- Pump maintenance
- User safety
- Pump specifications
- Alarms and error messages
- Alarm history
- Backlight
- Basal
- BG target
- Bolus delivery
- Bolus history
- Bolus units
- Bolus Wizard
- Carb ratios
- Carb units
- Daily totals
- Default screen
- Delivery accuracy
- Drive motor
- Dual Wave bolus
- Easy bolus
- Infusion pressure
- (insulin) sensitivity
- Insulin type
- Low resv (reservoir) warning
- Meter value
- Normal bolus
- Occlusion detection
- Percent temp basal
- Power supply
- Prime function
- Prime history
- Program safety checks
- Pump size
- Pump weight
- Remote control
- Reservoir
- Square Wave bolus
- Status screen
- Temporary (temp) basal rate
- Time and date screen
- Water tight
- Bolus Wizard specifications
- Default settings
- Icon table
- Menu map
Optimizing pump therapy 77
Deliver Easy Bolus
Practice using the Easy Bolus feature while looking at the pump screen as you count the beeps. After you are
familiar with Easy Bolus, you can use the audible tones for bolus delivery without having to look at the screen.
NOTE - Easy bolus only works from the HOME screen.
1. From the HOME screen, press . The SET EASY BOLUS screen
will appear. The single “step” value will be flashing.
2. Press the number of times needed for your bolus amount.
Watch the amount change on the screen with each press. When
your total bolus amount appears on the screen, press ACT.
The pump will vibrate or sound a different tone for each press.
Listen/feel to count the steps without looking at the screen.
For example: You need to deliver a 1.0 unit bolus, and your step size is set to 0.1.
Each time you press , the units increase by the “step” amount.
To deliver 1.0 units, you need to press the button 10 times. (10 X 0.1 = 1.0)
The screen will show 1.0 units.
NOTE - Pressing or
ESC
will cancel the Easy bolus.
Total # Steps = 4
Total # Button Presses = 4
4
3
2
1
0
1 Step = 0.5 units
1 Step = 0.5 units
1 Step = 0.5 units
1 Step = 0.5 units
1.00.5 1.5
Total # Units
2
Example:
2.0 unit Easy bolus with
a step amount set at 0.5 units
0
SET EASY BOLUS
0.1
U
(flashing)
MiniMed
SET EASY BOLUS
1.0
U
ACT
x 10