User Manual
Table Of Contents
- Table of contents
- Glossary
- Before you begin...
- The basics...
- Basic programming
- Starting on insulin
- Using Bolus Wizard
- Optimizing pump therapy
- Insulin pump therapy follow-up
- Utilities
- Troubleshooting and alarms
- Troubleshooting
- My pump has a no delivery alarm...
- What happens if I leave the battery out too long?
- Why doesn’t my pump battery last very long?
- What is a CHECK SETTINGS alarm?
- My screen appears distorted...
- I can’t get out of the priming loop...
- The pump is asking me to rewind...
- My bolus stopped...
- My pump buttons are not acting right during a bolus...
- My pump won’t display my BG reading from my meter...
- I dropped my pump
- I submerged my pump in water
- Alarms
- Alarm conditions
- Troubleshooting
- Pump maintenance
- User safety
- Pump specifications
- Alarms and error messages
- Alarm history
- Backlight
- Basal
- BG target
- Bolus delivery
- Bolus history
- Bolus units
- Bolus Wizard
- Carb ratios
- Carb units
- Daily totals
- Default screen
- Delivery accuracy
- Drive motor
- Dual Wave bolus
- Easy bolus
- Infusion pressure
- (insulin) sensitivity
- Insulin type
- Low resv (reservoir) warning
- Meter value
- Normal bolus
- Occlusion detection
- Percent temp basal
- Power supply
- Prime function
- Prime history
- Program safety checks
- Pump size
- Pump weight
- Remote control
- Reservoir
- Square Wave bolus
- Status screen
- Temporary (temp) basal rate
- Time and date screen
- Water tight
- Bolus Wizard specifications
- Default settings
- Icon table
- Menu map
76 Chapter 6
Easy bolus
The EASY BOLUS button allows a quick way to deliver a Normal bolus. You will pre-set the settings for this
feature in the EASY BOLUS OPTION screen in the BOLUS MENU. Your pump is sent from the factory with the Easy
Bolus feature set to on. If you do not want to use Easy Bolus, turn it off.
NOTE - If you are using the remote control, the Easy Bolus must be set to on.
After you setup Easy Bolus, with each press, you can increase the Normal bolus amount by a fixed amount,
called a “step.” Before you can deliver an Easy Bolus, you must set the amount in the EASY BOLUS ENTRY screen.
This amount equals the number of units of insulin for each step. The maximum number of steps is 20 and can be up
to your maximum bolus limit.
Once you set your step amount, you can program your Easy Bolus. When you are in the HOME screen, each time
you press the Easy Bolus amount increases by one “step.” You will hear a beep or feel a vibration for each step
increase. Each beep is a different tone. This makes it easy for you to count the beeps while you are programming
your Easy Bolus.
Easy bolus setup
1. Go to the EASY BOLUS option screen.
Main > Bolus > Easy Bolus
Select
On/Set
and press
ACT
. If you do not want
to use Easy Bolus, select
Off
and press
ACT
.
Step value setup
You can set the step value from 0.1 to 2.0 units (factory setting: 0.1). Set the step to a number that is convenient to
use and easy to multiply. The maximum number of steps is 20 up to the maximum bolus limit, whichever occurs
first.
2. In the EASY BOLUS ENTRY screen, the “step”
value will appear flashing. Change the value and
press ACT (The step value is the increment you
will use for your Easy bolus.)
3. The screen will return to the BOLUS MENU.
Your step amount is now programmed and Easy
Bolus is ready to use. Exit the menus.
MiniMed
EASY BOLUS OPTION
On/Set
Off
EASY BOLUS ENTRY
0.
5 U
/
EASY BOLUS ENTRY
0.
1 U
(flashing)
ACT
MiniMed MiniMed
BOLUS MENU
MiniMed
Set Bolus
Bolus History
Bolus Wizard Setup
ESC