User Manual
126 Chapter 12
Bolus Wizard specifications
notes:
■
Active insulin is based on timing of previous bolus insulin and insulin type.
Four to six hours after a bolus, the majority of insulin has been absorbed, but
a very small amount is active for a few more hours. The Bolus Wizard
automatically calculates this and subtracts the appropriate amount. The
details will appear in the ESTIMATE DETAILS screen during the bolus
programming steps.
food
carb ratio
food estimate
estimate = +
(current BG - BG Target)
insulin sensitivity
correction estimate
Food intake: 45gr
(Meter) BG: 160
Food: 3.0U
Correction: 2.0U
Active Ins: 1.0U
Est total: 4.0U
(values shown
are for
example only)
ESTIMATE DETAILS
ACT to proceed,
ESC to back up
mp6024875_012_a.book Page 126 Monday, June 23, 2003 10:54 AM