User Manual

Pump specifications 121
Drive motor
The pump motor has a unique, patented design with an integrated safety check system. The system delivers in
precise increments.
Dual Wave bolus
Delivers a Normal bolus followed by a Square Wave bolus (limited by maximum bolus).
Easy bolus
Features programming using audible tones (or vibrate pulses) in user-determined increments. Beep mode range: 0
to maximum bolus; Vibrate mode range: 0 to 20 steps or maximum bolus, whichever comes first.
Default step increment: 0.1 unit
Step size < maximum bolus.
Adjustable step size: 0.1 to 2.0 units per step.
Accessible from the remote control or pump buttons.
Infusion pressure
Maximum infusion pressure and occlusion pressure: 13.7 PSI.
(insulin) sensitivity
Maximum settings: 8
Factory default: 50 mg/dL (2.8 mmol/L)
Insulin type
Pump users can select one: U100 fast acting (default) U100 regular acting
Low resv (reservoir) warning
Values are based on displayed amount, not actual amount.
range: 10 250 mg/dL
(0.5 13.9 mmol/L)
warning limits: less than 20 or greater than 100 mg/dL
(less than 1.1 or greater than 5.6 mmol/L)
time: 2 24-hours, and
@1:00 hour before empty
08:00 hours (default when time is selected)
units: 5 50 units, and
@ 1/2 amount remaining
20 units (factory default)
mp6024875_012_a.book Page 121 Monday, June 23, 2003 10:54 AM