User Manual
Optimizing pump therapy 73
5. The SQUARE DURATION screen will appear. Enter the amount of time you want the
Square Wave bolus to last and press
ACT.
6. Press ACT to accept and deliver the bolus. The BOLUS DELIVERY screen will appear.
The pump will beep or vibrate at the start and end of the bolus. As the bolus is delivered,
the screen will show the bolus type and amount until the total units have been delivered.
The screen will then default to the HOME screen.
4. For a
Square Wave bolus:
The SET SQUARE BOLUS screen will appear.
Change the amount if desired. Press ACT to
accept.
For a Dual Wave bolus do these steps:
NOTE - The number of units you program for the
SET DUAL BOLUS TOTAL is a total of both the
Normal and Square Wave bolus units.
a. The SET DUAL BOLUS TOTAL
screen will appear. Change the
amount if desired. Press
ACT to
accept.
b. In the next screen, notice the
screen shows the Normal (Now)
and Square portions of the Dual
Wave bolus. Press
ACT to
accept the portions suggested by
Bolus Wizard, or press
/
to change these portions then
press
ACT.
MiniMed
SET SQUARE BOLUS
4.0U
Estimate
ACT
MiniMed
SET DUAL BOLUS TOTAL
4.0U
Estimate
ACT
MiniMed
Now:
0%
3.0U
Square:
100%
1.0U
ACT
MiniMed
SQUARE DURATION
Sq: 1.3 u
0:30
ACT
MiniMed
BOLUS DELIVERY
0.
1U
3:24P
NORMAL
ACT
mp6024875_012_a.book Page 73 Monday, June 23, 2003 10:54 AM