Using Bolus Wizard 57
3. The SET START TIME 2 screen will appear.
4. The screen will return to the EDIT SETTINGS screen. You can now program your BG targets.
BG targets
The BG target setting allows you to set your pre-meal blood glucose targets. Because this amount may vary
throughout the day, your pump allows you to set up to eight (8) BG targets. Your healthcare professional may only
have you program one or two BG targets when you first start using the Bolus Wizard feature.
NOTE - BG targets are normally between 90 – 140 mg/dL (or 5.0 – 7.8 mmol/L). If your targets are outside
this range, a warning message will appear on the screen.
1. In the EDIT SETTINGS screen,
select
BG Targets and press ACT.
2. Enter your first BG target and press
ACT.
NOTE - The start time for your first BG target is midnight (12:00A or 00:00) and cannot be changed.
To set just one ins sensitivity: To set more than one ins sensitivity:
Press ESC.
a. Set the start time for
ins sensitivity 2 and
press
ACT.
b. Set the value for
ins sensitivity 2 and
press
ACT.
c. Repeat steps a and b
to set more values.
d. Press
ESC when you are
done.
_:_
SET START TIME 2
MiniMed
(flashing)
mg/dL
per unit insulin
__
_:_
SET START TIME 2*
MiniMed
(flashing) set, then
press
ACT
__
per unit insulin
mg/dL
_:_
INS SENSITIVITY 2*
MiniMed
(flashing) set, then
press
ACT
__
per unit insulin
mg/dL
* Depending on how many
values you set, this num-
ber can be 2 through 8.
EDIT SETTINGS
MiniMed
Carb Ratios:
BG Targets:
--->
BG Units:
mg/dL
- - -
12:00A
SET TARGET BG 1
MiniMed
- -
mg/dL
first
BG target
(flashing)
ACT
mp6024875_012_a.book Page 57 Monday, June 23, 2003 10:54 AM