User's Manual

Table Of Contents
Warnings and precautions
84
Physicians should instruct patients to avoid any sharp blows or pressure directly
at the Pump location. A direct hit by an object such as a baseball may damage the
Pump and/or injure the Pump pocket. Extreme pressure on the Pump may cause
the sutures and the tissue pocket to be damaged. As a result, the Pump may move
from its pocket and possibly dislodge from the Catheter.
If the Pump has been damaged by such a blow, the patient should be instructed to
contact their physician. Physicians may order a pressure bandage or a reduction
in the patient’s physical activity until the pump pocket reheals completely.
Exposure to Extreme Body Temperatures
Exposure to elevated body temperatures in excess of 104°F (40°C) will
compromise the negative pressure reservoir safety feature of the Medtronic
MiniMed 2007D Implantable Pump. Glucose levels should be monitored closely
if this occurs.
Sterilization
The Pump, Catheter and the refill kit are sterilized with ethylene oxide (EtO).
The Pump System is NOT Reusable.
If the sterile package has been opened, damaged or tampered with DO NOT USE
the packaged device.
NOTE: Do Not Re-Sterilize the Implantable Insulin Pump
NOTE: Do Not Re-Sterilize the Side Port Catheter
NOTE: Do Not Re-Sterilize Components of the Refill Kit.
The Pipette (MMT-4104), may be resterilized using autoclave, EtO or according
to hospital procedures.
The Template (MMT-4106) may be resterilized using steam autoclave or EtO per
hospital procedure. This template is made from polycarbonate plastic and will
have a limited life in steam autoclave cycle. The template will not withstand dry
heat autoclave cycles.