User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Pump refill procedure
72
Perform the refill procedure
Fill out the refill form
Fill out the top part of the Refill Form (Appendix C) with the patient’s I.D., refill
date, insulin lot number, previous refill volumes, insulin medication remaining,
and the name of the person performing the refill procedure. This information can
be helpful for diagnostic purposes.
Label syringes
Use a sterile marker and Steri-Strips
®
to label one refill syringe “RB” (rinse
buffer syringe) and the other “IN” (insulin syringe).
Prepare the refill syringe for emptying the Pump
1. Firmly attach a stopcock to the RB syringe and
attach the 18 gauge needle to the stopcock.
Check connections to ensure they are secure.
Draw 5 ml of Rinse Buffer into the RB refill
syringe.
2. Expel all air bubbles from the refill syringe.
Remove the 18 gauge needle from the
stopcock.
3. Fill the hub of the refill needle with Rinse
Buffer from the RB refill syringe and attach it
to the stopcock. (See Figure 12) Prime the refill
needle completely. Close the stopcock.
4. Obtain the weight and record it on the Refill
Form. Set the syringe aside.
Figure 12: Filling the
hub of the refill needle