User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Pump refill procedure
70
Supplies and solutions
The refill kit (MMT-4105) is intended exclusively for use with the
Medtronic MiniMed 2007D Implantable Insulin Pump System. The refill kit con-
sists of a specially designed syringe and stopcock for removing fluids from and
placing fluids into the Pump. Use only refill needles (MMT-4102) available from
Medtronic MiniMed to perform all Pump refill procedures. This 22 gauge needle,
featuring a rounded, lubricated tip and a beveled side hole, has been specially
developed to mate with the Pump fill port.
Use of other needles may result in damage to the
Pump septum in the fill port, and may allow body
fluids to enter the Pump.
Below is a list of the supplies and solutions necessary to perform the refill proce-
dure:
Supplies:
• Steri-Strips
®
and Markers
• The Patient’s PPC, Glucose Monitoring Equipment, Drapes
• One (1) Scale (0.01 gram resolution)
• One (1) 250 ml Sterile Beaker
• One (1) Port Locating Template - MMT-4106
• Two (2) Refill Kits - MMT-4105 - Three (3) Medtronic MiniMed MMT-
4102 Refill Needles (Extra needles should be available for use.)
• Four (4) 18 Gauge Regular Bevel Needles
Solutions:
• One (1) 10 ml Vial, Rinse Buffer Solution, MMT-2008
• Two (2) 10 ml Vials, U-400 Insulin, stored at room temperature for a
minimum of 10 hours prior to surgery.
!
WARNING