User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Pump implantation
57
Figure 7: Venting the syringe head space
2. With the stopcock closed, insert the refill needle into the Pump fill port.
3. Open the stopcock. Maintain downward pressure on the barrel of the syringe
to ensure the inlet valve of the Pump remains open.
Never push on the refill syringe plunger to fill the
Pump. When the refill needle is properly seated in
the fill port of the Pump, the vacuum in the reser-
voir will draw the insulin from the syringe.
4. Allow the Pump vacuum to draw the INSULIN into the reservoir. When the
INSULIN stops moving, the Pump is filled. Close the stopcock and remove
the syringe.
5. Prepare the IN1 syringe for aspiration. Remove any air from the syringe,
prime the needle, close the stopcock and obtain a vacuum by pulling back on
the plunger until it locks. Press the lock into the plunger groove to be sure it
is firmly secured. A fluid barrier of at least 2 mL should remain in the
syringe.
!
WARNING