User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Personal Pump Communicator (PPC)
51
Technical history codes
In the “HISTORY” menu, the user can access clinical history from the PPC and
the Pump. Each number code corresponds to a particular event. The following
table lists each of the number codes with their corresponding events.
Technical history PPC codes
TELEMETRY COMM
ERROR
3
The PPC and Pump are not communicating. Reposition the PPC over the
Pump, then press SEL and ACT. If the error message persists, contact
Medtronic MiniMed.
TELEMETRY COMM
ERROR
20
Move the PPC away from any other Pump and perform the request again.
Display Screen Message Message Meaning or Action Required
Clinical Event Code Clinical Event Code
Version Error 1 Time Out 2
No Response 3 Retry Packet 4
No Synchronization 5 Bad CRC 6
Invalid transmission 7 RX Overflow 8
Invalid Op code 9 TX Underflow 10
Unknown IRQ Vector 12
Bolus Total Error 13 NMI Occurred 14
Bad Duplicates 15 Suspend Alarm 16
Stop Pump Alarm 17 Diagnostic Rate Alarm 18
Prime Pump Alarm 19 Exclusion List Full 20
Max Clock Stealer 21 Min Clock Stealer 22
Non Initialized PPC 23 Pump Self Test Error 24
Pump Initialized 25 Pump Reset to Defaults 26
Hourly Maximum 28