User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Personal Pump Communicator (PPC)
50
Personal Pump Communicator messages
Display Screen Message Message Meaning or Action Required
AUTO OFF
PUMP SUSPENDED
Auto Off time interval has elapsed. Pump operation is suspended.
BOLUS 0.0U A bolus has been programmed and is being delivered.
CHECK PUMP
STATUS
The PPC battery has been replaced. The PPC needs to check the Pump
status.
LOW RESERVOIR The PPC has recognized the Pump has 800 units (2 ml) or less insulin
remaining in its reservoir. Schedule a Pump refill as soon as possible.
Allowing the reservoir to completely empty may damage the Pump.
EMPTY RESERVOIR The PPC has recognized the Pump has 400 units (1 ml) or less insulin
remaining in its reservoir. Schedule a Pump refill as soon as possible.
Allowing the reservoir to completely empty may damage the Pump.
HOURLY MAX
EXCEEDED
You attempt to deliver more than 2.5 times the bolus maximum in one
hour. To clear the message, press SEL and ACT. You may exceed this
limit by programming another bolus within 10 minutes.
PPC LOW BATTERY The PPC battery is low. The PPC can be used for at least 24 hours.
Replace the AA battery immediately
PPC NOT INITIALIZED The PPC has not been initialized to a Pump.
PUMP LOW
BATTERY
The Pump battery energy is low but still functioning. Schedule a
replacement as soon as possible.
PUMP INITIALIZED The PPC has been “married” to a Pump. Press SEL and ACT.
PRIMING XX U A Priming Bolus has been programmed and is being delivered.
PUMP RESET The Pump was reprogrammed to the preset factory values.
PUMP SELF TEST FAIL A Pump malfunction was detected during a Self Test.
PUMP STOPPED The Pump is in stop mode.
PUMP STOPPED
1 or 2 or 3 or 4 or 5 or 6
When the Pump recognizes a system malfunction, it automatically stops
and insulin delivery ceases.
PUMP SUSPENDED The Pump is in suspend mode
PUMP ERROR
0 or 1
User attempts to initialize the PPC to a Pump that is not compatible with
it. Clear by pressing SEL and ACT. Verify the personal ID of the Pump
responding corresponds to the personal ID entered into the PPC.
PUMP ERROR
40 or 41
The Pump has invalid data. The PPC will not initialize to Pump.