Stop Pump
This screen allows the Physician to stop the operation of the Pump.
1. From the “PUMP REFILL” screen in the
Supervisor Menu, press SEL until the “STOP
PUMP” screen appears.
2. Press ACT and the word “CONFIRM” appears
highlighted with the entire screen flashing. Press
ACT again to confirm.
3. Place the PPC near the Pump and allow the
communication to complete.
4. The PPC will beep four times to confirm the
“STOP PUMP” command. To restart the Pump,
press SEL and “PUMP STOPPED” will start to
flash. Press ACT. Place the PPC near the Pump.
When the communication is complete, the PPC
returns to the Time/Date screen.
Supervisor password
If the password is lost, access the Supervisor Menu as follows:
Program the time to midnight (12:00 am or 00:00) and the date to
January 01.
Enter the password 0000 within one minute. If the password is not
entered within one minute the time will be reset to midnight again.
Record the new password in the patient’s chart.
Exit supervisor
To exit the Supervisor Programming Menu, press SEL until the “EXIT SUPER-
VISOR” screen appears, then press ACT.