User's Manual

Table Of Contents
Personal Pump Communicator (PPC)
47
6. The entire screen will be flashing. Press ACT
again to confirm the Diagnostic Delivery Rate.
7. Place the PPC near the Pump and complete the
communication process.
8. The words “DIAGNOSTIC RATE” appear
briefly on the screen.The PPC then returns to the
Time/Date screen.
NOTE: Press SEL and the PPC will indicate
“DIAGNOSTIC RATE.”
NOTE: Diagnostic Rate Delivery can be terminated at any time
by using the “SUSPEND PUMP” feature.
Initialize to factory defaults
This feature resets all programmed parameters in the Pump to preset factory
default values. To access this feature, the Pump must be in the “PUMP
STOPPED” mode.
1. From the “PUMP STOPPED” screen, press SEL.
The screen will begin flashing.
2. Press the and buttons at the same time until
the “ENTER SUPERVISOR PASSWORD
screen appears. Re-enter the Supervisor
Password, pressing ACT after each letter or
number is entered. The factory preset password
is: YIQ8.
10 U/H
CONFIRM
DIAGNOSTIC
RATE
PPC
COMMUNICATING
08:13 OCT 12
DIAGNOSTIC RATE
08:13 oct 12
PUMP STOPPED
ENTER
PASSWORD
SUPERVISOR
0000