User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Personal Pump Communicator (PPC)
46
Diagnostic rate
A diagnostic rate is intended for use only when the
Pump is filled with rinse buffer.
The Diagnostic Rate feature is used to help “diagnose” Pump delivery problems.
Fill the Pump with rinse buffer before using this feature. To access this feature,
the Pump must be in the “PUMP STOPPED” mode.
1. From the “PUMP STOPPED” screen, press SEL.
The screen will begin flashing.
2. Press the and buttons at the same time until
the “SET SUPERVISOR PASSWORD” screen
appears. Re-enter the supervisor password,
pressing ACT after each letter or number is
entered. The factory preset password is: YIQ8.
3. When the Supervisor Password is successfully
entered, the screen will change to “PUMP
REFILL”. Press SEL until the “DIAGNOSTIC
RATE” screen appears.
4. Press ACT and the word “SET” appears flashing.
Use the and buttons to enter a Diagnostic
Rate (units/hour).
5. Press ACT and the word “SET” changes to
“DELIVER”. “NO” also appears flashing. Use
the and buttons change “NO” to “YES”,
then press ACT.
!
WARNING
08:13 OCT 12
PUMP STOPPED
0000
SET
SUPERVISOR
PASSWORD
_ _ U/H
DIAGNOSTIC
RATE
10 U/H
SET
DIAGNOSTIC
RATE
10 U/H
DELIVER
DIAGNOSTIC
RATE NO