46
Diagnostic rate
A diagnostic rate is intended for use only when the
Pump is filled with rinse buffer.
The Diagnostic Rate feature is used to help “diagnose” Pump delivery problems.
Fill the Pump with rinse buffer before using this feature. To access this feature,
the Pump must be in the “PUMP STOPPED” mode.
1. From the “PUMP STOPPED” screen, press SEL.
The screen will begin flashing.
2. Press the and buttons at the same time until
the “SET SUPERVISOR PASSWORD” screen
appears. Re-enter the supervisor password,
pressing ACT after each letter or number is
entered. The factory preset password is: YIQ8.
3. When the Supervisor Password is successfully
entered, the screen will change to “PUMP
REFILL”. Press SEL until the “DIAGNOSTIC
RATE” screen appears.
4. Press ACT and the word “SET” appears flashing.
Use the and buttons to enter a Diagnostic
Rate (units/hour).
5. Press ACT and the word “SET” changes to
“DELIVER”. “NO” also appears flashing. Use
the and buttons change “NO” to “YES”,
then press ACT.
!
WARNING
08:13 OCT 12
PUMP STOPPED
0000
SET
SUPERVISOR
PASSWORD
_ _ U/H
DIAGNOSTIC
RATE
10 U/H
SET
DIAGNOSTIC
RATE
10 U/H
DELIVER
DIAGNOSTIC
RATE NO