User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Personal Pump Communicator (PPC)
44
6. The calculated accuracy will be displayed briefly.
Record this value on the refill form.
Priming
A priming bolus is not intended as a therapeutic bolus.
It should only be used when the Pump is not
implanted.
The Priming Bolus is used during the preparation of the Pump for implantation.
This feature “Primes” the Pump piston chamber, outlet port and Catheter. The
Pump will pulse approximately 500 times during this function. To access this fea-
ture, the Pump must be in “PUMP STOPPED” mode.
1. From the “PUMP STOPPED” screen, press SEL.
The screen will begin flashing.
2. Press the and buttons at the same time until
the “ENTER SUPERVISOR PASSWORD”
screen appears.
3. Re-enter the supervisor password, pressing ACT
after each letter or number is entered. The factory
preset password is: YIQ8. If the password has
been changed, enter the new password.
When the
Supervisor Password is successfully entered, the
screen will change to “PUMP REFILL”. Press
SEL until the “PRIMING” screen appears.
4. Press ACT and “NO” appears flashing. Press the
and buttons to change to “YES”, then press
ACT.
CALCULATED
ACCURACY
_ _ %
!
WARNING
08:13 Oct 12
PUMP STOPPED
ENTER
SUPERVISOR
PASSWORD
YIQ8
PRIMING
PRIMING NO
99.8U