User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Personal Pump Communicator (PPC)
41
Set an Audio Bolus from the main menu bolus screen
1. From the Time/Date screen press the button.
The PPC beeps either one or two times,
depending upon the audio bolus step amount set.
The user counts the number of beeps to
determine how much insulin was programmed.
2. In this example the PPC was programmed for a
0.8U increment, and the PPC beeps two times for
the 0.8U bolus. Press ACT and the audio
sequence repeats to confirm the bolus amount.
Press ACT twice and the Pump will be
programmed. The PPC beeps once when the
programming is completed. Place the PPC on the Pump to complete the
communication process.
NOTE: While an audio bolus is being delivered, the button
will not function.
Variable Bolus
To access the Variable Bolus feature enter “SETUP II” and press SEL one time.
1. The screen will display “VARIABLE BOLUS”.
Press ACT.
2. The screen will now display “SET VARIABLE
BOLUS” and “ON or OFF” will be flashing. Use
the and buttons to select “ON” or “OFF”,
then press ACT.
3. Place the PPC near the Pump to complete
communication. If “ON” was chosen the
Variable Bolus option is now available in the
Main Menu/Bolus screen.
NOTE: If “OFF” is chosen, the Variable
Bolus option (Square, Dual) will not be available in the
(Main Menu) Bolus screen.
SET BOLUS
0.8U
IMM
PPC
COMMUNICATING
VARIABLE
BOLUS
ON
SET
BOLUS
PPC
COMMUNICATING