User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Personal Pump Communicator (PPC)
40
Setup II
Audio Bolus
The Audio Bolus feature allows the user to deliver a bolus without looking at the
PPC screen. There are two Audio Bolus increments, 0.4U and 0.8U. With each
press of the button, the PPC will beep one time (0.4U setting) or two times
(0.8U setting), depending on the delivery amount selected per button press.
Activating the Audio Bolus feature
1. From the “SETUP II” screen, press ACT. The
“AUDIO BOLUS" screen is shown.
2. Press ACT, and the screen changes to “SET
AUDIO BOLUS” with “OFF/ON” flashing. Use
the and buttons to choose “ON” to activate
the Audio Bolus feature or “OFF” to disable it.
Then press ACT.
3. If “ON” was chosen, the “SET STEP AUDIO
BOLUS” appears with a step value of 0.4U or
0.8U flashing on the screen. Use the and
buttons to choose between a step rate of 0.4U or
0.8U. Press ACT. The Audio Bolus feature is
now activated.
NOTE: If “OFF” is chosen, pressing the
button will have no
effect.
AUDIO BOLUS
ON
0.4U
SET
ON
AUDIO BOLUS
0.4U
SET STEP
AUDIO BOLUS
0.4U