User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Personal Pump Communicator (PPC)
38
Alarms
This feature is previously described in this Chapter.
Self test
1. Self Test allows the user to perform a diagnostic
test of the Pump and PPC operating system.
Messages are relayed between the PPC and
Pump.
From the “SETUP PUMP” screen, press
ACT. Press SEL until the “SELF TEST” screen
is displayed. Press ACT.
2. Place the PPC near the Pump and complete the
communication process.
3. Verify that the following events occur:
• A series of tones will be heard from the Pump (4 beeps).
• An alarm tone will be heard from the PPC.
• The backlight on the PPC will turn on.
• The PPC screen will briefly activate all possible display icons, numbers,
etc.
• The PPC will vibrate.
NOTE: If the above test results do not occur or the screen
displays irregular characters, please notify
Medtronic MiniMed. If the PPC displays the Medtronic
MiniMed logo and software version, the PPC has
restarted due to a low battery. Replace the battery
immediately.
If all electronics “pass” the Self Test, the screen will automatically change to
“PPC PASSED/PUMP PASSED.” After several seconds, the screen will return to
the Time/Date screen.
SELF TEST
PPC
COMMUNICATING