User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Personal Pump Communicator (PPC)
37
Setup Pump
The “SETUP PUMP” screen permits access to the primary Setup menus for the
Pump. Setup features discussed previously in this chapter are referenced here.
Other Setup features not previously discussed are presented here. Press SEL until
the “SETUP PUMP” screen is displayed, then press ACT to access the SETUP
menus.
Time and Date
This feature is previously described in this Chapter.
Auto off
Auto Off is a safety feature, reminding the user to update insulin medication
delivery programming in the Pump. An alarm can be set after a period of pro-
gramming inactivity, from one to 16 hours. At the onset of the alarm the Pump
will automatically be placed in SUSPEND mode.
1. From the “SETUP PUMP” screen, press ACT.
Press SEL to reach the “AUTO OFF” screen,
then press ACT. The screen will display “AUTO
OFF” with flashing dashes/time. Use the and
buttons to select the number of hours before
an “AUTO OFF” alarm occurs.
2. Press ACT. In this example, a time duration of
10 hours was selected. The PPC will alarm if the
user did not attempt to program the Pump during
the previous 10 hours and be placed in
"SUSPEND" mode.
3. Place the PPC near the Pump and complete the
communication process.The PPC will beep once
and return to the Time/Date screen.
AUTO OFF
_ _ HRS
SET
AUTO OFF
10 HRS
PPC
COMMUNICATING