User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Personal Pump Communicator (PPC)
36
7. Press SEL and the screen will change to
“EST PUMP BATT”. This screen indicates the
Pump battery status, during no-load (battery
power is not used) and load (extended
communication sequence) conditions. The Pump
is set to alarm for low battery when the loaded
(LD) voltage is at or below 2.5 volts.
8. Press SEL and the screen will read
“EXIT HISTORY.” Press ACT. The PPC will
return to the Time/Date screen.
Jan 02
EST PUMP BATTERY
2.9V
NO LD LD
2.7V
EXIT HISTORY