- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Personal Pump Communicator (PPC)
34
Personal events
This feature allows the user to enter event codes into the PPC memory, and
record the time and date of entry. Preset event codes are: 1 = meal, 2 = snack, 3 =
sick and 4 = exercise. In addition, other event codes A, B and C can be entered to
record other important events. These other event codes should be documented
prior to their use.
To access the “EVENTS” screen in the main menu, “ON” must be activated in
the “SETUP II” menu, “PERSONAL EVENTS” screen.
1. From the Time/Date screen, press SEL until
“SETUP PUMP” is displayed, then press ACT.
Press SEL until “SETUP II” is displayed, then
press ACT. Press SEL until “PERSONAL
EVENTS” screen is displayed.
2. Press ACT and “ON” or “OFF” begins flashing.
3. Use the and buttons to select “ON”, then
press ACT again. The “PERSONAL EVENTS”
screen will now appear on the main menu. Allow
the PPC to return to the Time/Date screen.
4. To set an event: From the Time/Date screen press
SEL until the “EVENT” screen is displayed then
press ACT. The screen changes to “SET
EVENT” with the word “MEAL” flashing. Use
the and buttons to select the desired event.
5. Press ACT and the current time will appear
flashing. Use the and buttons to enter the
time the event occurred. Press ACT when the
proper time is displayed. Then allow the screen
return to Time/Date.
NOTE: Only historic or current event times can be entered.
NOTE: If the Personal Events feature is turned “OFF” in
“SETUP II”, events cannot be entered into the PPC.
OFF
PERSONAL
EVENTS
EVENTS
PERSONAL
ON/OFF
SET
PERSONAL
EVENTS
ON/OFF
SET
EVENT
MEAL
SET TIME
EVENT
MEAL
07:32Am