User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Personal Pump Communicator (PPC)
32
6. The screen will indicate
“PPC COMMUNICATING.” Place the PPC near
the Pump and complete the communication
process.
7. The screen will briefly display the calculated 24
hour basal dose based on the basal rate
programming. In this example a total of “8.4U”
will be delivered.
To set multiple basal profiles in the other patterns
(A,B,C), select the pattern in "SETUP II" menu
and follow the same procedure.
Program a temporary basal rate
A Temporary Basal Rate is often used when a brief change in basal delivery is
required, for example during exercise.
Set a temporary basal rate
1. From the time and date display press SEL until the
“TEMP BASAL” screen is displayed.
2. Press ACT and the “SET DURATION” screen
appears. The time duration of the Temporary Basal
Rate is displayed as flashing dashes. Press the
and buttons to enter a time duration
, in 30
minute increments from 30 minutes to 24 hours.
3. Press ACT and “SET AMOUNT” screen appears.
The amount of the Temporary Basal Rate is now
flashing. Press the and buttons to enter a
delivery rate
. Press ACT again.
PPC
COMMUNICATING
24 HOUR TOTAL
8.4U
TEMP BASAL
_ _ _ _ U/H
TEMP BASAL
_ _U/H00:30
SET DURATION
TEMP BASAL
1.5U/H00:30
SET AMOUNT