User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Personal Pump Communicator (PPC)
31
Setting basal rate profiles in each delivery pattern
Each of the basal delivery patterns is a set of up to 48 basal rates, one for each
half-hour of the day.
1. Press SEL until the “BASAL RATE” screen is
displayed. A basal pattern is selected (for
example Pattern A). Press ACT.
2. A “SET RATE” and “1” is now displayed to the
right of the “A” indicating this programming will
set the 1
st
basal rate within the “A” profile. The
“0.2U/H” is now flashing, indicating the value
can be changed. Use the and arrow keys to
program a new value, for example, “0.4 U/H”,
and then press ACT.
NOTE: 00:00 indicates a start time of MIDNIGHT in 24hr.
display mode. 12:00am indicates a start time of
MIDNIGHT in 12hr. display mode.
3. The screen now displays “SET TIME”, with a
time of “00:30” or “12:30am” flashing (24 or 12
hour respectively) and a “2.” This screen allows
the second basal rate to be set. Enter a start time
for the 2
nd
basal rate within the “A” profile, for
example “04:30.” Press ACT. (Example: a second basal rate of 0.4U/H
starting at 04:30.)
4. This screen changes to “SET RATE” again,
indicating the 2
nd
basal rate can now be
programmed. Use the and arrow buttons to
enter a new rate, for example “0.2U/H”, and then
press ACT.
5. A “3” now appears on the screen with “SET
TIME.” Follow the same procedure previously
described and program a new profile. If no
additional profiles are needed change the
flashing time to dashes (by pressing ) and press
ACT.
BASAL RATE: A NOW
00:00 0.2U/H
BASAL RATE: A
SET RATE
1
00:00 0.2U/H
SET TIME
04:30 0.4U/H
BASAL RATE: A 2
SET RATE
BASAL RATE: 2
04:30 0.4U/H
SET TIME
05:00
BASAL RATE: 3
0.2U/H