User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Personal Pump Communicator (PPC)
30
Programming a basal rate
Basic basal rate programming was described earlier in this chapter. This section
describes additional basal rate options.
Programming basal delivery pattern
The PPC allows three basal delivery patterns. One such basal pattern could be
used for a working day, another for a weekend day, etc. Each of the basal delivery
patterns is a set of up to 48 basal rates, one for each half-hour of the day. Pattern
A is the factory pre-set. To access profiles B or C you must enter the
“SETUP PUMP” screens.
1. Press SEL until the “SETUP PUMP” screen is
displayed. Press ACT. Press SEL again to access
the basal rate profile screen, “DELIVERY
PATTERN”.
2. Press ACT and the screen will change to “SET
DELIVERY PATTERN”. Use the and keys
select the pattern preferred; A, B, or C. Each
pattern can contain up to 48 different basal rates.
Press ACT after choosing a pattern.
3. Place the PPC near the Pump and allow the
communication to complete.
NOTE: When the PPC times out, press SEL until the
“BASAL RATE” screen is displayed. The basal pattern
selected in "SETUP PUMP" will now appear to the right
of “BASAL RATE” A, B, or C.
DELIVERY
PATTERN
A
SET
DELIVERY A,B,C
PATTERN
PPC
COMMUNICATING