Personal Pump Communicator (PPC)
29
Suspend mode
The Suspend Pump mode allows the user to cancel a bolus delivery, while still
delivering a basal rate of 0.2 U/hr.
1. From the Time/Date screen, press SEL until the
“SUSPEND PUMP” screen is displayed. Press
ACT. The screen will display a flashing
“SUSPEND PUMP” message. Press ACT again.
2. Place the PPC near the Pump and complete the
communication process.
3. When the communication is complete, the Pump
will beep 3 times and the PPC screen will change
to “PUMP SUSPENDED”. All four segments of
the insulin delivery icon are shown. The PPC
will beep every half-hour as long as the Pump
remains suspended.
NOTE: To restart the Pump, press SEL. The “PUMP
SUSPENDED” screen will begin flashing. Then press
ACT.
SUSPEND PUMP
PPC
COMMUNICATING
08:13
PUMP SUSPENDED
OCT 12