User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Personal Pump Communicator (PPC)
29
Suspend mode
The Suspend Pump mode allows the user to cancel a bolus delivery, while still
delivering a basal rate of 0.2 U/hr.
1. From the Time/Date screen, press SEL until the
“SUSPEND PUMP” screen is displayed. Press
ACT. The screen will display a flashing
“SUSPEND PUMP” message. Press ACT again.
2. Place the PPC near the Pump and complete the
communication process.
3. When the communication is complete, the Pump
will beep 3 times and the PPC screen will change
to “PUMP SUSPENDED”. All four segments of
the insulin delivery icon are shown. The PPC
will beep every half-hour as long as the Pump
remains suspended.
NOTE: To restart the Pump, press SEL. The “PUMP
SUSPENDED” screen will begin flashing. Then press
ACT.
SUSPEND PUMP
PPC
COMMUNICATING
08:13
PUMP SUSPENDED
OCT 12