User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Personal Pump Communicator (PPC)
26
Set a Square Wave Bolus
A Square Wave Bolus of insulin is delivered evenly over a preset period of time,
from 30 minutes to 4 hours. A Square Wave Bolus may be desirable when eating
long meals such as at banquets or receptions, high fat meals, or to compensate for
gastroparesis. During a Square Wave Bolus, the programmed basal rate is deliv-
ered simultaneously to the bolus.
To access this feature you must first turn the Variable Bolus feature “ON” in the
“SETUP II” menu.
1. From the Time/Date screen, press SEL. The
“BOLUS” screen is displayed, showing the last
bolus programmed with the time and date
flashing.
2. Press ACT and the “SET BOLUS TYPE” screen
appears. Press the and buttons to select
“SQUARE”. Press ACT.
3. The “BOLUS” screen appears, with dashes under
“EXT” flashing. Use the and buttons to
enter an extended
bolus amount. Press ACT.
4. Blinking dashes will now appear under the bolus
amount entered. Use the and buttons to
enter a time duration
for the Square Wave Bolus,
in one-half hour increments from 30 minutes to 4
hours.
5. Press ACT and the “CONFIRM BOLUS” screen
is displayed, with the screen flashing. Confirm
the Square Wave Bolus by pressing ACT again.
6. Place the PPC near the Pump and complete the
communication process.
08:13
BOLUS
PROG
IMM
10U _ _U
--
Jan 02
EXT
SET
SQUARE
BOLUS TYPE
SET BOLUS
IMM
_ _U 4.0U
EXT
SET BOLUS
IMM
2:00
EXT
_ _U 4.0U
CONFIRM BOLUS
IMM EXT
2:00
_ _U 4.0U
PPC
COMMUNICATING