User's Manual

Table Of Contents
Personal Pump Communicator (PPC)
25
Set a Normal bolus with the Variable Bolus feature turned
on
1. From the Time/Date screen press SEL until the
“BOLUS” screen is displayed. The last bolus
value programmed and the Time and Date will
be flashing.
2. Press ACT and the “SET BOLUS TYPE” screen
appears. If “NORMAL” is not flashing, use the
and buttons to select “NORMAL.” Press
ACT.
3. The “SET BOLUS” screen appears, with dashes
under “IMM” flashing. Use the and buttons
to enter an immediate
bolus amount.
4. Press ACT and the “CONFIRM” screen is
displayed, with the screen flashing. Confirm the
bolus amount by pressing ACT again.
5. Place the PPC near the Pump and complete the
communication process.
6. When the bolus programming is complete, the
PPC will beep once and then briefly display the
amount of insulin currently delivered.
The Pump will beep at each of the first five
strokes (if audio feedback is ON). The PPC
beeps at the end of the bolus. Three segments of
the insulin delivery icon will be displayed and
spinning slowly during the bolus delivery. By
pressing SEL you can read the amount of insulin
delivered.
08:13
BOLUS
IMM
_ _ U _ _ U
Jan 02
EXT
SET
NORMAL
BOLUS TYPE
SET BOLUS
IMM
_ _ U
CONFIRM
IMM
2.6U
PPC
COMMUNICATING
08:23
BOLUS 5.0U
JAN 04