Set a Normal bolus with the Variable Bolus feature turned
off
1. From the Time/Date screen, press SEL.
The “BOLUS” screen is displayed, with the time
and date flashing.
2. Press ACT and the “SET BOLUS” screen
appears. The dashes under “IMM” are flashing.
Press the and buttons to enter an immediate
bolus amount.
3. Press ACT and the “CONFIRM” screen is
displayed, with the screen flashing. Confirm the
bolus amount by pressing ACT again.
4. Place the PPC near the Pump and complete the
communication process.
5. When the bolus programming is complete, the
PPC will beep once and then briefly display the
amount of insulin currently delivered.
The Pump will beep at each of the first five
strokes (if audio feedback is ON). The PPC
beeps and at the end of the bolus. Three
segments of the insulin delivery icon will be
displayed and spinning slowly during the bolus
delivery. By pressing SEL you can read the
amount of insulin delivered.
08:13
BOLUS
PROG
IMM
10U 10U
Jas2
--
EXT
SET BOLUS
IMM
_ _ U
CONFIRM
2.6U
IMM
PPC
COMMUNICATING
08:13
BOLUS 0.0U
Jan 02