User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Personal Pump Communicator (PPC)
22
Program a basal rate
1. From the Time/Date screen, press SEL until the
“BASAL RATE” screen is displayed. Preset
delivery pattern “A”, a basal rate of 0.2 U/H, and
the word “NOW” is flashing. Press ACT.
2. A “1” now appears to the right of the “A”
indicating that this programming will effect the
1
st
basal change within the “A” pattern, (there
are 3 patterns available [A, B, C] and 48 basal
changes possible within each pattern). The
flashing 0.2 U/H indicates the value can be
changed. Use the and buttons to change the
value and then press ACT.
NOTE: 00:00 indicates a start time of MIDNIGHT in 24hr.
display mode. In 12hr. display mode, the screen indicates
the start time as 12:00am.
3. The screen now displays “SET TIME”, and a
time of 00:30 or 12:30 am (24 or 12 hour
respectively) and a “2.”
If one basal is all that will be used, press ACT
two times. If more than one basal rate is to be
programmed, enter a start time and amount of the new basal rate for that time
period, then press ACT and enter the new basal rate. The user can enter a
new basal rate at 30 minute intervals, up to 48 basal rates.
4. Place the PPC near the Pump and complete the
communication process.
5. The PPC will briefly display the calculated total
basal dose for 24 hours, based on the values and
times entered in the Basal Rate programming
screen. In this example, the total basal dose is
4.8 U/day.
08:26 OCT 12
BASAL RATE: A NOW 1
00:00 0.2U/H
BASAL RATE A
00:30 _ _ U/H
s
2
PPC
COMMUNICATING
24 HOUR TOTAL
4.8U