User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Personal Pump Communicator (PPC)
21
4. Place the PPC near the Pump and complete the
communication process. The PPC screen will
automatically advance to the “PERSONAL ID”
screen. Press ACT.
5. The first of the 32 possible ID locations is
flashing. Enter the patient ID (alpha-numeric) by
pressing the and buttons and then ACT
after each entry. Continue to press ACT,
activating each “0” until the screen changes.
6. Place the PPC near the Pump and complete the
communication process.
7. Press SEL until the “SET SUPERVISOR
PASSWORD” screen is displayed. Then press
ACT.
8. The screen now reads, “SET SUPERVISOR
PASSWORD”. Press ACT. Use the and
buttons to enter a new supervisor password
(alphanumeric), pressing ACT after each entry.
9. Record the password in the patient’s chart and
implant form.
PPC
COMMUNICATING
SET PERSONAL ID
• • • • • • • • • • • • • • • •
• • • • • • • • • 20KOLO5
PPC
COMMUNICATING
SET
SUPERVISOR
PASSWORD
YIQ8
SET
SUPERVISOR
PASSWORD
0000