User's Manual

Table Of Contents
Personal Pump Communicator (PPC)
20
7. Place the PPC near the Pump and complete the
communication process. The PPC screen will
return to the “PERSONAL EVENTS” screen.
Allow the PPC to time out and return to the
Time/Date screen.
NOTE: Adding screens to the main menu, such as “PERSONAL
EVENTS” increases the number of SEL button presses
required to reach “SETUP PUMP.”
Lock maximum bolus/basal, enter personal ID and password,
stop Pump
To access the Supervisor Mode press SEL until the “SETUP PUMP” screen is
displayed. Then press and hold down the and buttons simultaneously until
the “ENTER SUPERVISOR PASSWORD” screen appears.
Patients should not be given the Supervisor
Mode password, to avoid the accidental pro-
gramming of a large priming bolus (99.8 U) or
diagnostic insulin rate.
1. The first zero will be flashing. Press the and
buttons to select the first digit, then press ACT.
The screen advances to the second zero. Press
the and buttons to select the second digit,
then press ACT. Repeat for the last two digits.
The factory pre-set password is YIQ8.
2. Entry into the Supervisor Mode is indicated by
the screen “PUMP REFILL.”
3. Press SEL until the “SET MAXIMUMS" screen
is displayed, and then ACT to reach “SET
MAXIMUMS”.
Press the and buttons to select “ON” if the
patient is not
given access to this feature, or
“OFF” if the patient is
permitted access. Press
ACT again.
PPC
COMMUNICATING
!
WARNING
ENTER
SUPERVISOR
PASSWORD
0000
PUMP
REFILL
SET
MAXIMUMS
OFF/ON