User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Personal Pump Communicator (PPC)
19
Set maximum bolus, basal rate and time display format
This programming is performed in the “SETUP II” menus. These screens allow
healthcare professionals to limit the maximum amount of insulin a patient can
deliver, either when taking a bolus or setting a new basal rate. Access to the
“SETUP II” menus is through the “SETUP PUMP” screen.
1. Press SEL until the “SETUP PUMP” screen is
displayed. Press ACT and press SEL to reach
the “SETUP II” screen. Press ACT and then
SEL to reach the “MAX BOLUS” screen.
2. Press ACT and the maximum bolus amount
(units) will start flashing. Press the and
buttons to change the maximum allowable bolus
(0.2 to 25.0 units) and then press ACT again.
3. Place the PPC near the Pump and complete the
communication process. The PPC screen will
automatically change to the “MAX BASAL”
screen.
4. Press ACT and the screen will change to “SET
MAX BASAL RATE”. The maximum basal
amount will start flashing. Press the and
buttons to change the maximum allowable basal
rate (0.2 to 35.0 units/hour) and then press ACT
again.
5. Place the PPC near the Pump and complete the
communication process. The PPC screen will
automatically change to “TIME FORMAT.”
6. Press ACT and the screen will change to “SET
TIME FORMAT.” Press the and buttons to
select either a 12 hour (AM/PM) or 24 hour
(military time) format, and then press ACT.
MAX BOLUS
_ _ U
MAX BOLUS
SET
_ _ U
PPC
COMMUNICATING
MAX BASAL RATE
SET
0.2U/H
PPC
COMMUNICATING
SET
TIME FORMAT
12/24 HOUR