User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Personal Pump Communicator (PPC)
17
Set the time and date
The time and date settings must be correct to ensure accurate calculation of insu-
lin delivery, daily totals, and the proper display of insulin activity history.
1. Press SEL until the “SETUP PUMP” screen is
displayed, then press ACT two times. The first
two digits of the time (hours) will be flashing.
Use the and buttons to select the correct
hour, then press ACT. The last two digits of the
time (minutes) will be flashing. Use the and
buttons to select the correct minute, then press
ACT. Repeat the programming process to enter
information for the year, month and day.
2. After completing the programming process, then
place the PPC near the Pump. The PPC will
display “PPC COMMUNICATING” while
transferring the time and date information to the
Pump. The PPC will then move to the next
screen, “AUTO-OFF.” Skip the “AUTO-OFF”
screen by pressing SEL once to reach the next
screen, “ALARMS”.
08:32
SET
TIME-DATE
2004
Jan 02
PPC
COMMUNICATING