User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
Personal Pump Communicator (PPC)
15
Part 1: PPC/Pump system initialization
The Implantable Insulin Pump arrives from Medtronic MiniMed with preset fac-
tory default values. During the initialization process these preset values are
downloaded into the PPC memory. The preset values can then be changed by the
healthcare professional, allowing the system to be personalized for each patient.
The factory default values are as follows:
Initialize the PPC
When the healthcare professional receives a new PPC it must be “married” to the
Pump. Following are the basic steps used to initialize a Pump System the day
prior to implant:
1. The PPC is delivered without a battery in place.
After installing a new battery, the PPC will beep
six times and the screen will identify the PPC
software used (see Chapter 3, Install/Replace the
Main Battery).
2. The screen now changes to
“PPC NOT INITIALIZED”, and the PPC will
beep six times once every minute until the
initialization process is started. Press SEL and
then ACT, then quickly place the PPC over the
Pump.
Bolus delivery type: Normal Locked set maximums: Off
Maximum bolus: 25.0 U Password: YIQ8
Audio bolus: Off Personal Events status: OFF
Audio feedback: Disabled Personal ID: 000 (32 characters)
Auto Off duration: Off PPC alarm type: High
Basal delivery pattern: A Refill amount: 25 g
Maximum basal rate: 35.0 U/H Time format: 12 hours
Insulin concentration: 400 U/ml Variable Bolus status: Off
PUMP
PPC 8 102 100
— — —