User's Manual
Table Of Contents
- Table of contents
- List of figures
- CHAPTER 1 Description
- CHAPTER 2 Indications and contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- CHAPTER 4 Pump implantation
- Preprogramming and pre-testing the Pump
- CHAPTER 5 Pump refill procedure
- CHAPTER 6 Explanting the Pump System
- CHAPTER 7 Warnings and precautions
- CHAPTER 8 Adverse reactions
- CHAPTER 9 System alarms and messages
- Pump alarms
- Alarm feedback
- Pump low battery
- Depleted pump battery
- System error
- Pump self test fail
- PPC low battery
- PPC alarms
- Low reservoir
- Empty reservoir
- Telemetry communication error
- Initialize alarm
- PPC not initialized
- Battery replacement
- Initialize to factory defaults
- Pump stopped
- Pump suspended
- Auto off
- Hourly maximum exceeded
- Pump alarm table
- Pump alarms
- CHAPTER 10 Troubleshooting Pump System under-delivery
- CHAPTER 11 Technical specifications
- APPENDIX A Label information symbol dictionary
- APPENDIX B Implant worksheet
- APPENDIX C Refill form
- APPENDIX D Precautions and general procedures
- APPENDIX E Pump rinse procedure
- APPENDIX F Side Port Catheter flush procedure
- Supplies and solutions
- Preparing for the procedure
- Flushing the Side Port Catheter
- Program minimal basal rate
- Remove insulin and fill with rinse buffer
- Equilibrate and pull rinse buffer through system
- Flush side port catheter
- Remove rinse buffer and fill with insulin
- Equilibrate and pull insulin through system
- Remove guide needles and record refill amount
- Program new basal rate
- Remove rinse buffer from catheter
- APPENDIX G Stroke volume measurement
138
• One (1) sterile stopcock - 3 way
Solutions
• One (1) 10 mL vial, Rinse Buffer Solution, MMT-2008
Preparing for the procedure
NOTE: Before beginning any refill, flush, stroke volume
measurement, carefully read Appendix D, Precautions
and General Procedures, and keep this in mind as you
perform each procedure.
Record patient’s blood glucose
Record the patient’s blood glucose value at the start of the procedure. Monitor
blood glucose every 30 minutes, or as needed.
Measuring stroke volume
1. In a sterile field before the procedure; prepare the apparatus. Prime the
stopcock, pipette, and needle.
2. Turn the stopcock as shown in Figure 18. Press on the syringe to obtain a
liquid meniscus in the pipette.
Figure 18: Stroke volume measurement setup