User's Manual

Table Of Contents
133
NOTE: This is the only time a plunger should be manually
pushed down.
Remove rinse buffer and fill with insulin
The RINSE BUFFER is removed from the Pump fluid system and the system is
filled with insulin. This procedure uses the following volumes in syringes 3, 5,
and 6:
Syringe 3 with residual RINSE BUFFER from the previous step
Syringe 5 with 20 ml INSULIN
Syringe 6 with 30 ml INSULIN
Syringe 3: Rinse Buffer - Residual
1. Prime the needle. Close the stopcock.
2. Obtain a vacuum by pulling back on the plunger until it locks. Press the lock
into the plunger groove and be sure it is firmly secured.
3. Enter the Pump with the refill needle and syringe.
4. Open the stopcock and remove the rinse buffer. After the fluid level stops
moving, wait an additional 30 seconds to make sure all of the RINSE
BUFFER and air is removed.
5. When the time has elapsed, close the stopcock, remove the syringe, and
discard it.
Syringe 5: Insulin - Degassed
1. Prime the needle and close the stopcock.
2. Enter the Pump fill port with the refill needle.
3. Open the stopcock and allow the Pump to fill completely with INSULIN.
When the fluid level stops moving, the Pump is filled.
4. Close the stopcock and remove the syringe.
5. Prepare syringe 6 for aspiration. Obtain a vacuum by pulling back on the
plunger until it locks. Press the lock into the plunger groove to be sure it is
firmly secured.
6. Re-enter the Pump.